Copyright notice: This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each author's copyright. These works may not be reposted without the explicit permission of the copyright holder.


Abstract

We have developed a system architecture that will allow a surgeon to employ direct hand-eye coordination to conduct medical procedures in a remote microscopic environment. In this system, a scaled real-time video image of the workspace of a small robotic arm, taken from a surgical microscope camera, is visually superimposed on the natural workspace of a surgeon via a half-silvered mirror. The robot arm holds a small tool, such as a microsurgical needle holder or microsurgical forceps, and the surgeon grasps a second tool connected to a position encoder, in this case, a second robot arm. The views of the local and remote environments are superimposed such that the tools in the local and remote environments are visually merged. The position encoder and small robot arm are linked such that movement of the tool by the operator produces scaled-down movement by the small robot tool. To the surgeon, it seems that his hands and the tool he or she is holding are moving and interacting with the remote environment, which is really microscopic and at a distance. Our current work focuses on using a position-controlled master-slave robot linkage with two 3 degree of freedom haptic devices, and we are pursuing the use of a 6-to-7 degree of freedom master-slave linkage to produce more realistic interaction.


Notes

Associated centers: VASC, MRTC, SRI, and FRC
Associated labs/groups: Neurobotics Laboratory, Human-Robot Interaction Group, and Visualization and Image Analysis
Associated project: Sonic FlashlightTM

Number of pages: 8


Text Reference

S. Clanton, D.C. Wang, Y. Matsuoka, D.M. Shelton, and G.D. Stetten, "A Novel Machine Interface for Scaled Telesurgery," SPIE Medical Imaging, Vol. 5367, February, 2004, pp. 697 - 704.


BibTeX Reference

@inproceedings{Clanton_2004_5090,
   author = "Samuel Clanton and David C Wang and Yoky Matsuoka and Damion Michael Shelton and George D Stetten",
   title = "A Novel Machine Interface for Scaled Telesurgery",
   booktitle = "SPIE Medical Imaging",
   month = "February",
   year = "2004",
   volume = "5367",
   pages = "697 - 704"
}


The Robotics Institute is part of the School of Computer Science, Carnegie Mellon University.
For updates and comments, please see these instructions.
This page maintained by robotwebmaster@ri.cmu.edu