W. Ang, C. Riviere, and P. Khosla
Proc. IEEE/RSJ Intl. Conf. Intelligent Robots and Systems, Vol. 2, October, 2001, pp. 1106-1111.

Jump to: Download | Abstract | Notes | Text Reference | BibTeX Reference

Download [Help]

Adobe portable document format (pdf) [463 KB]

Copyright notice: This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each author's copyright. These works may not be reposted without the explicit permission of the copyright holder.

Abstract

This paper presents the development and initial experimental results of the first prototype of Micron, an active hand-held instrument to sense and compensate physiological tremor and other unwanted movement during vitreoretinal microsurgery. The instrument incorporates six inertial sensors, allowing the motion of the tip to be computed. The motion captured is processed to discriminate between desired and undesired components of motion. Tremor canceling is implemented via the weighted-frequency Fourier linear combiner (WFLC) algorithm, and compensation of non-tremorous error via a neural-network technique is being investigated. The instrument tip is attached to a three-degree-of-freedom parallel manipulator with piezoelectric actuation. The actuators move the tool tip in opposition to the tremor, thereby suppressing the erroneous motion. Motion canceling experiments with oscillatory motions in the frequency band of physiological tremor show that Micron is able to reduce error amplitude by 45.3% in 1-D tests and 37.2% in 3-D tests.

Notes

Sponsor: National Institutes of Health
Grant ID: R21 RR13383

Associated center: MRTC
Associated lab/group: Medical Instrumentation Lab
Associated project: Micron: Intelligent Microsurgical Instruments

Text Reference

W. Ang, C. Riviere, and P. Khosla, "Design and implementation of active error canceling in hand-held microsurgical instrument," Proc. IEEE/RSJ Intl. Conf. Intelligent Robots and Systems, Vol. 2, October, 2001, pp. 1106-1111.

BibTeX Reference

@inproceedings{Ang_2001_4094,
   author = "Wei-Tech Ang and Cameron Riviere and Pradeep Khosla",
   title = "Design and implementation of active error canceling in hand-held microsurgical instrument",
   booktitle = "Proc. IEEE/RSJ Intl. Conf. Intelligent Robots and Systems",
   month = "October",
   year = "2001",
   volume = "2",
   pages = "1106-1111"
}