Carnegie Mellon Robotics Institute
Wei-Tech Ang, Cameron Riviere, and Pradeep Khosla
Proceedings of the Third International Conference on Medical Image Computing and Computer-Assisted Intervention, October, 2000, pp. 878-886.
| Download |
|
| Abstract |
| This paper presents the first prototype of an active hand-held instrument to sense and compensate physiological tremor and other unwanted movement during vitreoretinal microsurgery. The instrument incorporates six inertial sensors (three accelerometers and three rate gyros) to detect motion of the handle. The movement of the instrument tip in three dimensions is then obtained using appropriate kinematic calculations. The motion captured is processed to discriminate between desired and undesired components of motion. Tremor canceling will be implemented via the weighted-frequency Fourier linear combiner (WFLC) algorithm, and compensation of non-tremorous erroneous motion via an experimental neural-network technique. The instrument tip is attached to a three-degree-of-freedom parallel manipulator, actuated by three piezoelectric stacks. The actuators move the tool tip in opposition to the motion of the tremor or other erroneous motion, thereby suppressing the error. Experimental results show that the prototype is able to follow one-dimensional and three-dimensional trajectories with rms error of 2.5 microns and 11.2 microns respectively. |
| Keywords |
| tremor, microsurgery, accuracy, medical robotics |
| Notes |
Sponsor: Johnson & Johnson Focused Giving Program Associated Center(s) / Consortia:
Medical Robotics Technology Center Associated Lab(s) / Group(s):
Surgical Mechatronics Laboratory Associated Project(s):
Micron: Intelligent Microsurgical Instruments and ASAP Number of pages: 9 |
| Text Reference |
| Wei-Tech Ang, Cameron Riviere, and Pradeep Khosla, "An Active Hand-held Instrument for Enhanced Microsurgical Accuracy," Proceedings of the Third International Conference on Medical Image Computing and Computer-Assisted Intervention, October, 2000, pp. 878-886. |
| BibTeX Reference |
|
@inproceedings{Ang_2000_3511, author = "Wei-Tech Ang and Cameron Riviere and Pradeep Khosla", editor = "Scott L. Delp, Anthony M. DiGioia, Branislav Jaramaz", title = "An Active Hand-held Instrument for Enhanced Microsurgical Accuracy", booktitle = "Proceedings of the Third International Conference on Medical Image Computing and Computer-Assisted Intervention", pages = "878-886", publisher = "Springer", month = "October", year = "2000", } |
| The Robotics Institute is part of the School of Computer Science, Carnegie Mellon University. Contact Us | Update Instructions |